Features
Sealing Features:
- Seal Lengths: 18″, 24″, 30″, 36″, 48″ (Twin head, 4 vacuum nozzles)
- Seal Width: 5/16″
- Bi-active jaws (top and bottom heated bars) with selector switch permitting heat from either or both jaws
- Pneumatically operated sealing jaws for high pressure sealing
- Independent digital seal and cool timers
- Digital sealing pressure switch with adjustable band to set high and low tolerances
- Seal Only switch to disable vacuum and retract the nozzle for use as an impulse sealer
Temperature Control Features:
- Temperature controller with digital readout to 350° F, accurate to +/- 3° F (high temp optional)
- Dual thermocouple sensors monitor heated jaws selected by operator (Top, Bottom, or Both)
- Adjustable temperature band for high and low alarm tolerances
- Seal time begins when set temperature is achieved to prevent inconsistencies with heat build up in sealing jaws, ensuring repeatable seals
Vacuum Gas Flush Features:
- Venturi Vacuum pump for quiet operation and fast evacuation of air (1/4 hp and hp electric vacuum pumps optional)
- Dual Vacuum/Gas Flush Nozzles (snorkels) to accommodate narrow or wide bags
- Digital timers for each stage of vacuum and gas flush (when applicable), up to 10 seconds
- Up to 2 stages of vacuum and 2 stages of gas flush
MedPAC Medical Control Features:
- ISO 11607 Compliant
- Fully calibratable pressure switch and temperature controller
- External ports in control cabinet for verification and calibration of internal controllers
- Real time seal process parameter monitoring (Temperature, Pressure, Time)
- Alarm function prevents operation when there is a deviation from validated settings
Safety Features:
- Pneumatic low-pressure jaw close with release if obstruction is detected
- Jaw close time out with release when delay is detected
- Heavy duty seal transformer(s) with thermal overload protection
- Emergency stop button on control panel
Versatility:
- Stainless Steel frame and cabinet for ease of cleaning and sterilization
- Locking Control Box cover to prevent unauthorized access to adjust settings
- Ported Exhaust for plumbing exhaust outside of a cleanroom to prevent contamination
- Adjustable sealing head position for vertical to horizontal use (Crank adjustable height optional)
- Designed to effectively package medical devices in a wide variety of bag materials
Specs:
- Foot switch cycle activation (Dual palm control optional)
- Air Requirement (min): 100 psi
- Electrical Requirement: 115v standard (230v optional)
- 3 year limited warranty
- Made in USA